Mitech MJB-W500B Micro - control series impact tes

Brand:

Mitech

Model:

Measuring range:

800mm

Resolution Accuracy:

±0.1°

Communication storage:

Working time:

Main unit weight:

450kg

Dimensions:

2124×600×1340mm

Package dimensions:

100*100*150cm, 120*100*160cm

Total weight:

450kg, 700kg

Product

  • Overview
  • Graphic
  • Technical
  • Configuration
  • Related

Product

  • Overview
  • Graphic
  • Technical
  • Configuration
  • Related
  • Product Overview

    Mitech MJB-W micro-control series impact testing machine, throuth the computer control pendulum movement, for metal, which can be for metal, non-metallic and composite materials under dynamic load resistance to the impact of the test. It uses the computer control, with stable performance and high degree of automation, integrated operation. Its structure is strong, simple structure, high efficiency. The machine is widely used in steel, steel pipe and other toughness of ferrous metals and alloys of its anti-pendulum impact performance of the accurate measurement,It is the necessary professional precision testing equipment for improving production efficiency and saving production costs.

    Function & application

    It is widely used in the accurate determination of pendulum impact resistance of ferrous metals and alloy materials, such as steel, steel pipe, etc..

    Working Principle

    The hemispherical punches of the impact testing machine impact and pass through the specimen at a certain speed and measure the energy consumed by the punch to evaluate the impact resistance of the specimen.

    Working Conditions

    l Operation Temperature:Ambient temperature45℃;

    l Relative humidity:20%~80%;

    l In an environment free from vibration,corrosive medium and high magnetic field

    l Installed on a flat basis.

    l Power supply voltage fluctuation does not exceed 10% of rated voltage.

    Instrument Features

    l Widely used in metallic and non-metallic processing industry, quality testing in quality inspection departments , scientific 

    research in higher education institutions and other fields;

    l It adopts single-chip control,and tilt pendulum, hanging pendulum, impact, are all under the computer control, and automatic 

    tilt pendulum, with high efficiency, especially for continuous impact test;

    l Zero adjustment, calibration, save, etc. without any analog adjustment link, the control circuit is highly integrated;

    l The system can realize the automatic calibration of the indication accuracy;

    l At the end of the test, the test data is automatically saved for later retrieval.

    l With features of novel shape, strong structure, high reliability, simple operation;

    l Can be directly observed in the display, intuitive readings, high test efficiency;

    l Using a single support column structure, cantilevered hanging pendulum, U-shaped pendulum hammer, with stable 

    performance and high precision;

    l Impact knife with screw mounting fixed, easy to replace;

    l Sample simple beam support can quickly replace the sample, easy to operate;

    l Take the brand computer, Chinese and English Windows operating platform, menu prompts, mouse operation, with fast 

    running, clear interface, simple operation and so on, to meet the needs of different materials test;

    l Real-time print output test data, user-friendly data processing, analysis and so on;

    l Equipped with safety protection pin, safety protection network, etc., to ensure safety test;

    l Consistent with GB, ISO, ASTM, DIN and other relevant domestic and foreign standards.

    Operating Method and Attentions

    Instrument Maintenance

    l Before using the instrument, please read the instruction manual carefully, understand the operation steps and precautions,to 

    avoid the damage caused by improper operation or personal safety accident;

    l Testing machine is a large precision instrument, should pay attention to avoid water, moisture. Exposed workbench and 

    attached parts should be coated with anti-rust oil to prevent rust;

    l It is forbidden to put samples before the pendulum hangs. Before placing the sample, make sure that the pendulum has been 

    stalled and that the insurance pin has been ejected;

    l Should pay attention to check the use of jaw support, pendulum, etc. are reliable to live tight to prevent the test results 

    inaccurate or accident occurred;

    l Pendulum shaft bearings at both ends of the factory has been refueling, users do not have to refuel, if it is repaired,then add 

    1-2 drops of sewing machine oil or watch oil, the rest of the power bearing plus Vaseline or butter;

    l If idle for a long time, at least once a week to prevent rust;

    l After the test, the pendulum to be placed in the vertical position, and the handle putter placed in the brake position;

    l Dont disassemble the instrument without authorization, maintenance related matters please contact MITECH after-sale 

    service department with 4000600280. 

  • Impact energy 250J、500J
    Impact speed 5.4m/s
    Tilt angle 150°
    Bearing span 40mm
    Angle of support on Sample bearing
    Angle of impact tool 30±1°
    Impact knife thickness 16mm
    Bearing fillet radius R1~1.5mm
    Impact blade radius R2~2.5mm
    Angular accuracy ±0.1°
    Pendulum torque 80.3848N.m、160.7695N.m
    Pendulum rotation center to the impact point (test center) distance 800mm
    Sample size 10×10(7.5、5、2.5)×55mm
    Power 180W
    Power supply 380V
    Dimensions (mm) 2124×600×1340mm
    Total Weight 450kg